Carbon dioxide monitor

Performance

  • ETCO₂

    • Measurement range: 0∼150 mmHg

    • Resolution: 1 mmHg

    • Accuracy:

      • 0∼40 mmHg:±2 mmHg

      • 41∼70 mmHg:±5% of readings

      • 71∼100 mmHg:±8% of readings

      • 101∼150 mmHg:±10% of readings

  • Respiration rate

    • Measurement range: 2∼120 rpm

    • Resolution: 1 rpm

    • Accuracy:

      • ±1 rpm

  • Safety characteristics

    • Degree of waterproof and dustproof: IPX1

    • Type of protection against electric shock: internally powered equipment

    • Degree of protection against electric shock: type BF applied part

  • Other parameters

    • Preheat time: 2 min

    • Response time: <3 s

    • Working hours: 8 h

    • Gas sampling rate: 60 ml/min∼100 ml/min

    • Alarm sound pressure level: 50 dB

      Physical Characteristics

      Dimension159 mm × 72 mm × 31 mm
      Weight270 g
      Gross weight of single piece472 g
      Inner box size202 mm (L) × 142 mm (W) × 90 mm (H)
      Outer box size43.2 cm (L) × 38.5 cm (W) × 24 cm (H)
      Quantity10 pieces
      Gross weight6.2 kg

      Operating Environment

      • Temperature
        5 °C ~ 40 °C

      • Relative Humidity
        30% ~ 75%, no condensation

      • Atmospheric Pressure
        700 hPa ~ 1060 hPa

      • Power Supply
        DC 3.7 V (Rechargeable lithium batteries)

      • Input Power
        ≤ 1.5 VA

      Accessories

      • User Manual
        1 book

      • Sampling Tube
        1 piece

      • USB Cable
        1 piece